Broadley Technologies 14 Manuel d'instructions Page 30

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 56
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 29
Model 14 Oxygen Analyzer
5-9
AFFECT - Conc. To set the choice of analog output to the
- nA concentration or temperature measurement
- ˚C/˚F
TYPE - 0/20 Choice of the analog output type
- 4/20
MODE - Lin Choice between a linear or dual range (see drawing below)
- Dual
LOWER XXXX Bottom scale value
MIDD. XXXX Mid-scale value (only in
dual mode)
UPPER XXXX Top of the scale value
SPECIAL PROG.
MODE - last Characteristics of the analog output during calibration,
- preset alarm system, maintenance or timer active cycles. Display
- live and output will be last stored value, a preset value, or a live
measurement
VALUE XX Preset value ( 0 to 21 mA)
TEST
Test the analog outputs in 1 mA increments (0 to 21 mA)
Output 1/2
Special Prog.
Test
Dual
Linear
4 0
20 20
16 15
12 10
8 5
0 102030405060708090100(%)
(mA)
UPPER
MIDDLE
LOWER
Vue de la page 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 55 56

Commentaires sur ces manuels

Pas de commentaire