Model 14 Oxygen Analyzer
Display Display in concentration units or % saturation
Display of the direct cell current in nA
Display of the sample temperature in °C/°F
Programming via menus
Units nA, ppm-mg/l, °C, °F, % saturation
Calibration Electrical zero, chemical zero, slope calibration in the
air, slope process calibration by comparison with a
laboratory measurement
Analog Outputs 2 x 0/4 - 20 mA isolated from input signal, 800 ohms
load maximum
- Measurement or temperature
- Mode : linear, bi-linear
- Accuracy : 0.1 mA Alarms - Number : 4
- Functions : alarm - system alarm - timer
- Hysteresis : 0 - 10%
- Delay : 0 - 999 seconds
- Breaking power : 250 VAC, 3A maximum
30 VDC, 0.5A maximum
Temperature Compensation Automatic in the range of 0 - 80 °C
RS485 (option) Speed : 300 - 9600 bauds
Insulation : galvanic
Station number : 32 maximum
Transmitter Protection IP 65 and NEMA 4X
Error Reports Cell current > 999 nA
Sample temperature > 80 °C
Ambient temperature > 80 °C
Slope calibration error
Zero calibration error (offset)
Immunity Against Electromagnetic Interferences EN 50082-2 and EN 50082-1
Electromagnetic Emission EN 50081-1 and EN 50081-2
Low Voltage Standard IEC61010-1
MATERIALS
Working Electrode Cathode : platinum
Counter Electrode Anode : silver
Membrane Holder PPS & silicone
Membrane PTFE
Transmitter Epoxy coated aluminum
Probe Body Stainless steel 316L
MAI
NTE
NANCE
Cleaning Clean the instrument with a soft tissue.
DO NOT use any aggressive agent.
1-3
Transmitter
Electromagnetic Compatibility
Materials
Maintenanace