Model 14 Oxygen Analyzer
Chapter2:
Description of the Analyzer
2.1 Transmitter
2.1.1 Presentation of the Transmitter
2-1
The electronic unit amplifies the signal of the amperometric measuring cell and converts it into a direct digital
readout in ppm, mg/l or % saturation. The transmitter is comprised of the following items:
Potentiostat which maintains the working electrode potential constant
Amperometric measuring module
Analog multiplexer
Microprocessor unit
The analog multiplexer allows measurement inputs from the measuring cell, temperature sensor and internal
checkpoints. Furthermore, the microprocessor operates the relays, the RS485 interface (optional) and the ana-
log outputs.
The unit has an internal concentration auto-ranging feature and a microprocessor-operated calibration routine.
The output of the potentiostat is monitored for possible overdriving of the potentiostat-output stage. This condi-
tion can occur when the connections to the measuring cell are due to an inoperable electrode or a defective ref-
erence electrode.